PTC Professional Temperature Calibrator

PTC (Professional Temperature Calibrator)

The PTC Series is a professional-grade temperature calibrator designed for both laboratory and field use. It offers high accuracy, fast heating and cooling, and a wide temperature range suitable for industries such as pharmaceuticals, energy, food processing, and general manufacturing. The calibrators are portable, robust, and built to provide reliable and consistent temperature calibration for various sensor types.

Features

  • Active dual-zone temperature control for improved temperature uniformity in the calibration well

  • Wide temperature range depending on model, covering approximately –90 °C up to 660 °C

  •  High accuracy when used with an external reference sensor, with very stable temperature control

  • Fast heating and cooling for reduced calibration cycle times

  • User-friendly interface with a full-color display and intuitive menu navigation

  • Portable and durable design suitable for field or industrial environments

  • Supports multiple sensor types, including RTDs and thermocouples

  • Compatible with calibration software and offers connectivity options for data logging and documentation

Specifications

Category

Specification

Temperature range

- 90 to 125°C/-130 to 257°F - PTC 125

-25 to 155°C / -13 to 311°F - PTC 155

33 to 350°C / 91 to 662°F – PTC 350

33 to 425°C/91 to 797°F – PTC 425

33 to 660°C/91 to 1220°F – PTC 660

Accuracy

±0.07°C to ±0.50°C (depends on sensor type) PTC 125/155/350/425/660

Stability

± 0.03°C / ± 0.054°F – PTC 125

± 0.013°C / ± 0.018°F– PTC 155

± 0.02°C / ± 0.036°F– PTC 350/425

± 0.04°C / ± 0.072°F – PTC 660

Resolution (user-selectable)

1°, 0.1°, 0.01°, or 0.001° - PTC 125/155/350/425/660

Heating time

13 - 15minutes (Depending on Temperature Ranges) – PTC 125

4 - 12minutes (Depending on Temperature Ranges) – PTC 155

7 minutes (Depending on Temperature Ranges) – PTC 350

10 minutes (Depending on Temperature Ranges) – PTC 425

20 minutes (Depending on Temperature Ranges) – PTC 660

Cooling time

30 – 75 mins (Depending on Temperature Ranges) – PTC 125

10 – 15 mins (Depending on Temperature Ranges) – PTC 155

12 mins (Depending on Temperature Ranges) – PTC 350

16 – 25 mins (Depending on Temperature Ranges) – PTC 425

15 – 36 mins (Depending on Temperature Ranges) – PTC 660

Time to stability

10 minutes – PTC 125/155/350/425/660

Immersion depth

190 mm / 6.3 in – PTC 125

160 mm / 6.3 in – PTC 155

140 mm / 5.5 in – PTC 350

150 mm / 5.91 in – PTC 425/660